Application 5

Measurement of NORM (scales) from the oil- and gas-industry

Relevant measurements of scales in tubings, valves and other contaminated components in the oil- and gas-industry can be made using portable gamma-ray spectrometers with NaI(Tl) or other room-temperature scintillation detectors, Material from wet or dried lagoons containing “produced water” from drilling rigs can also be surveyed. Spectra are analysed on-line with SODIGAM yielding nuclide-specific contamination information that will serve for immediate and well-founded ad-hoc decisions how to proceed.
Using specific batch-files one can precisely and fast analyse peaks from any nuclide in the spectrum.
An example of a measurement of scales in tubings from a Libyan oil rig is shown below.

Figure:  Spectrum from scales containing almost no thorium

Using fully automatic batch-file operation, several regions of the spectrum containing relevant peaks from Uranium and Thorium are analysed using SODIGAM.

This sample does not contain significant amounts of thorium. From the most clearly distinguished Th-peak at 2614 keV (approx. channel 880), the Th-contents is calculated as only 35.7± 5.0 Bq/kg. 

Moreover one does not find Uranium in the scales spectrum!! but only peaks from its progeny after 226Ra. The position of the strongest “uranium” peak at 1001 keV (this peak actually comes from 234mPa) is indicated in the spectrum with the brown marker line. .

The result of the quantitative SODIGAM spectrum analysis and nuclide assignment is shown below:

The activity of 222Rn and progeny is somewhat lower than the 226Ra activity because part of the noble gas 222Rn is released in the decay of 226Ra by alpha-recoil and due to its long halflife it can emanate from the sample.

Despite the short measuring time of only 622 seconds the SODIGAM analysis of the scales spectrum yields remarkably precise results which agree within uncertainties with HPGe spectrometry of the same sample.

Quotients of analysed results from 222Rn / 226Ra activities measured with different gamma-ray spectrometers are as follows:               HPGe detector:  0.9118 ± 0.0807      NaI(Tl) detector:  0.9451 ± 0.0960
i.e., around 6% to 9% of radon progeny emanate from the scales material. 